Organic Acid Analyser
Organic Acid Analyser based on an Ion Chromatography (IC) system is a sophisticated analytical tool used to detect and quantify organic acids. IC has been employed for ion analyses in solution samples obtained from various fields, such as semiconductor cleaning solutions, bodily fluids, and food, as well as for water quality control of river water or drinking water. The analyte ions in IC are mainly the common ions (e.g., Cl–, NO3–) found in environmental water or drinking water samples, but have also extended to weak acids, such as carboxylic acids, silicate and fluoride ions, owing to the wider range of targeted samples. However, the low resolution among weak acidic anions has been an important drawback of conventional IC and the instrumentation for it has undergone significant changes.

FEATURES
- Specifically designed for organic acid analysis
- Integrated tabletop systems with automatic injection system
- Partial loop injection
- Automatic calibration
- Uses Ion exclusion mode (+ reversed phase mode) column
- Highly selective detection with post column method
- Full control by PC
- Powerful data analysis software
SPECIFICATION
Conductivity Measuring Range |
up to 20000 µS/cm at 1 – 20 Hz, sample frequency, Auto zero function for the entire measuring range, Noise: 0.1 nS/cm |
Suppression | Two electrolytic Suppressor |
Column Oven Temperature | 30-90 ˚C |
Pump | Isocratic Pump (analytical, non-metal – Peek version), Pressure Range: 0-40 MPa (0-6000 PSI), Flowrate: 0.001 – 10 ml/min |
Degasser | Integrated 1-channel vacuum degasser |
Injection System | Automatic dual needle design to avoid system blockage with wash program and Programmable Injection Volume: 0.1 – 999.9 µ in increments of 0.1 µl, Sample Capacity: 120 samples (1.5 ml), Sample Loop: 100 µl, Carry Over: < 0.05 % |
Including |
|
INDUSTRIES
METHODS
ASTM D5542 | Standard Test Methods for Trace Anions in High Purity Water by Ion Chromatography |
ASTM D1982 | Standard Test Method for Titer of Fatty Acids |
ISO 10304 | Water quality — Determination of dissolved anions by liquid chromatography of ions |
SPARES & ACCESSORIES

Porous Solvent Filter

Injection Needle for Autosampler

Check Valve Cartridge/Ceramic

Secondary Sealing Ring Analytical

Primary Sealing Ring PE Analytical

Injection Port incl. Sealing PEEK Autosampler

PEEK Capillary 0.25*1.59 mm (1/16”) 3m

Finger Tight Fitting PEEK

Cationic Guard Column

Ferrule f.1/16″ Capillary Mat. PTFE TM used in IC Systems for Cond. Detector

Fitting 1/4-28 Gray with Knurled Screw Head 1/16”